林宗建,陆家瑜,吕成奇,于丽丽,瞿晓辉,滕新亚,邹德荣.4D下颌运动轨迹描记仪用于关系转移的精确性初探[J].口腔材料器械杂志,2019,28(1):23-29.
4D下颌运动轨迹描记仪用于关系转移的精确性初探
The study of four-dimensional mandibular motion trajectory measurement in oral clinical application
投稿时间:2018-09-11  修订日期:2018-12-18
DOI:10.11752/j.kqcl.2019.01.05
中文关键词:  4D下颌运动轨迹描记仪  完全可调式(牙合)架  下颌运动
英文关键词:4DTM Jaw Motion  System Mechanical articulator  Mandibular movement
基金项目:国家自然科学基金项目(编号:81470714,81870806,81800935);上海市浦江计划(编号:17PJD030)
作者单位E-mail
林宗建 上海交通大学附属第六人民医院口腔科, 上海 200233  
陆家瑜 上海交通大学附属第六人民医院口腔科, 上海 200233  
吕成奇 上海交通大学附属第六人民医院口腔科, 上海 200233  
于丽丽 上海交通大学附属第六人民医院口腔科, 上海 200233  
瞿晓辉 上海交通大学附属第六人民医院口腔科, 上海 200233  
滕新亚 上海交通大学附属第六人民医院口腔科, 上海 200233  
邹德荣 上海交通大学附属第六人民医院口腔科, 上海 200233 drzou@sjtu.edu.cn 
摘要点击次数: 1343
全文下载次数: 1104
中文摘要:
      目的 利用4D下颌运动轨迹描记仪(4DTM Jaw Motion System)记录下颌运动轨迹,测量计算切导斜度、髁导斜度等相关参数,初步探讨其获得的参数值用于(牙合)关系转移的精确性和可靠性。方法 选取个别正常(牙合)、无颞下颌关节疾患等符合入选标准的30名志愿者作为受试对象,男女各15例,平均年龄在25~35岁之间,利用4D下颌运动轨迹描记仪测量下颌运动的左右侧Bennett角、左右侧前伸髁导斜度以及前伸切导斜度等5组参数,同时运用传统机械式(牙合)架测量法计算相同参数,测量3次,取平均值。所得数据采用t检验进行统计学处理。结果 4D下颌运动轨迹描记仪和机械式(牙合)架测量所得参数平均值分别为: ①右侧Bennett角:(24.81±5.23) °和(25.35±4.05) °。②左侧Bennett角:(24.20±8.12) °和(26.58±4.98) °。③右侧前伸髁导斜度:(37.60±9.11) °和(34.13±6.74) °。左侧前伸髁导斜度:(38.25±10.32) °和(35.60±8.40) °。⑤前伸切导斜度:(39.97±9.43) °和(41.14±11.05) °。经过统计学分析,测量结果均在正常值范围内,两者之间差异无统计学意义(P>0.05)。结论 运用4D下颌运动轨迹描记仪测量(牙合)架相关参数所得数值具有精确性和可重复性,操作简便,舒适度好。
英文摘要:
      Objective The purpose of this research is to apply 4DTM Jaw Motion System for recording the track of mandibular movement and measuring relevant paraments, including incisal guidance inclination and condylar guidance inclination, so as to explore its precision and reliability. Methods Thirty subjects without malocclusion and TMJDS were selected for this experiment. We aimed to measure 5 paraments including Bennett angle, incisor inclination and condylar inclination, taking advantage of 4DTM Jaw Motion System and traditional articulator for the comparison of mean difference. Result All subjects were successfully operated and the paraments obtained are as follows:① The right Bennett angle for 4DTM Jaw Motion System and articulator are (24.81±5.23)° and (25.35±4.05)°, and the left are (24.20±8.12)° and (26.58±4.98)°. ② The right condylar inclination angle are (37.60±9.11)° and (34.13±6.74)°, simultaneously the left are (38.25±10.32)° and (35.60±8.40)°. ③ The incisor inclination are (39.97±9.43)° and (41.14±11.05)°, respectively. Statistical analysis showed there was no significant difference between digital and conventional methods. Conclusion 4DTM Jaw Motion System can be used to record and analyze the mandibular movement and has already shown its precision and reproducibility of transfer values. This measurement is more convenient and simpler for operation, and patients feel comfortable as well.